humidifier
1The environment requirement:
temperature 10℃--30℃
the relative humility ≤80%
environment pressure 86KPa—106KPa
2 work pressure
≤20KPa.
3 the electrical requirement
The power:AC220V±22V,50HZ
Fuse:0,5A
Power<85VA
4 The airway temperature test
Sensor type: temperature sensor
The test range:20℃--50℃
The error:±1℃
5 weight and size
W:135mm H:145mm D:95mm without chamber
1.2kg without chamber
2.0kg with chamber and water
6 the safety function
It is adaptive for IEC60601-1 1995
When the temperature inside chamber is over 95℃, the machine stops to heat.
When the temperature of airway is over 38℃, the machine will stop to heat and at the same time it will produce sound and display alarm signal. When the temperature of airway is low than 32℃,the machine will produce sound and display alarm signal too.
1The environment requirement:
temperature 10℃--30℃
the relative humility ≤80%
environment pressure 86KPa—106KPa
2 work pressure
≤20KPa.
3 the electrical requirement
The power:AC220V±22V,50HZ
Fuse:0,5A
Power<85VA
4 The airway temperature test
Sensor type: temperature sensor
The test range:20℃--50℃
The error:±1℃
5 weight and size
W:135mm H:145mm D:95mm without chamber
1.2kg without chamber
2.0kg with chamber and water
6 the safety function
It is adaptive for IEC60601-1 1995
When the temperature inside chamber is over 95℃, the machine stops to heat.
When the temperature of airway is over 38℃, the machine will stop to heat and at the same time it will produce sound and display alarm signal. When the temperature of airway is low than 32℃,the machine will produce sound and display alarm signal too.